Topcon GPT-2003 Manuale Utente Pagina 30

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 168
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 29
3-4
3 ANGLE MEASUREMENT
3.5 Repetition Angle Measurement
Repetition angle measurement can be done by horizontal angle right measurement mode.
Make sure the mode is Horizontal Angle Right measurement.
Operating procedure Operation Display
1
Press the [F4](
) key to get the function on page 2.
[F4]
2
Press the [F2](REP)key.
[F2]
3
Press the [F3](YES) key.
[F3]
4
Collimate the target A and press the [F1] (0SET)
key.
Collimate A
[F1]
5
Press the [F3] (YES) key.
[F3]
6
Collimate the target B using the horizontal clamp
and tangent screw.
Press the [F4](HOLD) key.
Collimate B
[F4]
7
Recollimate target A using the horizontal clamp
and tangent screw, and press the [F3](REL)key.
Collimate A
[F3]
8
Recollimate target B using the horizontal clamp
and tangent screw, and press the [F4](HOLD) key.
Collimate B
[F4]
9 Repeat 7 to 8 to measure the desired number of
repetitions.
[Example] 4 measurement
V : 90°10'20"
HR: 170°30'20"
0SET HOLD HSET P1
TILT REP V% P2
REPETITION ANGLE
> OK?
--- --- [YES][NO]
REP-ANGLE COUNT[ 0]
Ht: 0°00'00"
Hm:
0SET V/H REL HOLD
REPETITION ANGLE
INITIALIZE
> OK?
--- --- [YES][NO]
REP-ANGLE COUNT[ 0]
Ht: 0°00'00"
Hm:
0SET V/H REL HOLD
REP-ANGLE COUNT[ 1]
Ht: 45°10'00"
Hm: 45°10'00"
0SET V/H REL HOLD
REP-ANGLE COUNT[ 1]
Ht: 45°10'00"
Hm: 45°10'00"
0SET V/H REL HOLD
REP-ANGLE COUNT[ 2]
Ht: 90°20'00"
Hm: 45°10'00"
0SET V/H REL HOLD
REP-ANGLE COUNT[ 4]
Ht: 180°40'00"
Hm: 45°10'00"
0SET V/H REL HOLD
Vedere la pagina 29
1 2 ... 25 26 27 28 29 30 31 32 33 34 35 ... 167 168

Commenti su questo manuale

Nessun commento